Mindray BeneHeart D6 Defibrillator / Monitor
Mindray BeneHeart D6 defibrillator/monitor with 1-360 J biphasic defibrillation, an 8.4-inch display and 3/5/12-lead ECG monitoring for emergency and critical care.
ICEN supplies selected professional medical equipment. Product availability and market authorization vary by country.
A fully featured defibrillator/monitor for emergency and critical care
The Mindray BeneHeart D6 adds monitoring depth to the BeneHeart defibrillator platform. Manual defibrillation covers 21 selectable settings from 1 J to 360 J with a biphasic truncated exponential waveform and automatic impedance compensation, and charges to 200 J in less than three seconds or 360 J in less than seven seconds. Synchronized cardioversion delivers energy within 60 ms of the QRS peak.
Monitoring is the main difference from the D3: an 8.4-inch TFT colour display at 800 x 600 shows up to four waveform channels, and the unit monitors 3/5/12-lead ECG with respiration, SpO2, NIBP, EtCO2 (both Sidestream and Microstream), two IBP channels and two-channel temperature. A thermal recorder is built in and prints on 50 mm or 80 mm paper.
Noninvasive pacing runs from 30 to 210 ppm at 0 to 200 mA. Two 15.1 V / 5600 mAh lithium-ion packs can be fitted, giving roughly 400 shocks at 360 J, 12 hours of monitoring or 9 hours of pacing. Data leaves the unit over 5G or 2.4G Wi-Fi using IHE HL7 with DHCP, and the D6 integrates with a central monitoring system and the Pre-Hospital Emergency Information System. ICEN supplies the BeneHeart D6 to hospitals, emergency services and distributors, and confirms configuration and market requirements before quotation.
Applications
- Emergency department and resuscitation
- Pre-hospital emergency information systems
- Intensive and critical care transport
- Cardioversion and paced rhythm management
Who is this product for?
- Hospital emergency and critical care departments
- Emergency medical services with advanced monitoring needs
- Facilities needing 12-lead ECG at the bedside
- Distributors supplying advanced defibrillator/monitors
Selected Product Highlights
Specifications below are presented from the current ICEN product catalog. Confirm final configuration and technical documents with our team before purchase.
4-in-1: manual defibrillation, AED, pacing and monitoring
1-360 J, biphasic truncated exponential with impedance compensation
8.4-inch TFT colour display, 800 x 600, up to 4 waveform channels
3/5/12-lead ECG, SpO2, NIBP, EtCO2 (Sidestream and Microstream), 2-channel IBP, 2-channel temperature
Dual battery bays: about 400 shocks at 360 J, 12 h monitoring, 9 h pacing
Built-in thermal recorder (50/80 mm paper); IHE HL7 over 5G/2.4G Wi-Fi
Mindray BeneHeart D6 Defibrillator / Monitor - key specifications
| Brand | Mindray |
|---|---|
| Model | BeneHeart D6 |
| Category | Defibrillator / Monitor; professional biphasic defibrillator-monitor; 4-in-1 integrated design (manual defibrillation, AED, pacer, and monitoring modes) |
| Defibrillation energy | Manual mode output energy: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 70, 100, 150, 170, 200, 300, 360 J (21 selectable energy settings); up to 360 J escalating dose energy |
| Waveform | Biphasic truncated exponential waveform, with impedance compensation |
| Impedance compensation | Yes - automatic; voltage and energy delivery cycle are adjusted automatically to correspond with impedance. Patient impedance range (external defibrillation): 25 to 300 Ohm |
| Charge time | Less than 3 seconds to 200 J; less than 7 seconds to 360 J (with a new, fully charged battery). Energy accuracy: +/-2 J or 15% of setting, whichever is greater, into 50 Ohm |
| Synchronized cardioversion | Energy transfer begins within 60 ms of the QRS peak; within 25 ms of the External Sync Pulse |
| Power-on / ECG recovery time | Power on time less than 2 seconds with a new, fully charged battery; ECG recovery time less than 2.5 seconds |
| Display | 8.4 inch TFT Color LCD; 800 x 600 pixels; max 4 channels of waveforms; wave viewing time max 16 s (ECG) |
| Waveform channels | Display: max 4 channels; Recorder: max 4 channels |
| Monitoring parameters | 3/5/12-lead ECG with respiration, SpO2, NIBP, EtCO2 (Sidestream CO2 and Microstream CO2), IBP (2 channels) and 2-channel temperature (T1, T2, TD). SpO2 options: Mindray / Masimo / Nellcor. |
| AED mode | Automatic rhythm analysis with voice and text prompts. AED shock series energy level: 100 to 360 J, configurable; shock series: 1, 2, 3, configurable; default configuration meets 2015 AHA Guidelines; CPR mode with 1-channel ECG monitoring; sensitivity and specificity meets AAMI DF-80 |
| Pacing | Noninvasive pacing. Waveform: monophasic square wave pulse; pulse width 20 ms or 40 ms (+/-5%); refractory period 200 to 300 ms, +/-3% (function of rate); modes: demand or fixed; rate 30 ppm to 210 ppm (+/-1.5%); output 0 mA to 200 mA (+/-5% or 5 mA, whichever is greater); 4:1 pacing function |
| Alarms / clinical decision support | Arrhythmia detection: Yes; Alarms: Yes. AED automatic rhythm analysis; Glasgow 12-lead ECG report; Pre-Hospital Emergency Information System (PHEIS) supporting remote diagnosis; Adult/Paediatric mode automatically changes default energy, monitoring range and alarm limits |
| Data / event storage | Patient profiles: max 100 patients; Events: up to 1000 events per patient; Waveform storage: up to 24 hours of consecutive ECG waveform; Tabular trends: max 72 h of all measured parameters (resolution 1 min); Voice recording: max 180 min in total, max 60 min per patient; Data export to PC via USB flash memory |
| Connectivity | Transmits data through 5G/2.4G WiFi; supports international standard IHE HL7 protocol and DHCP (automatic IP address); USB flash memory data export; integrates with Central Monitoring System (CMS) and PHEIS |
| Battery | Type: 15.1 V, 5600 mAh rechargeable lithium ion battery pack; number: max 2; charge time less than 3 hours to 90% and less than 4 hours to 100% with equipment power off; 5-segment LED capacity indicator; capacity (two new, fully charged batteries): monitoring mode 12 hours (5-lead ECG, Resp, SpO2, 2-channel temp, CO2 and NIBP at 15-minute interval, Wi-Fi disabled), defib mode 400 times at 360 J discharge at intervals of 1 minute without recording, pacing mode 9 hours (50 Ohm load, 80 bpm, 60 mA). Battery package weight 0.57 kg each |
| Power supply | AC power: line voltage 100 to 240 V~ (+/-10%), current 1.8 to 0.8 A, frequency 50/60 Hz (+/-3 Hz). DC power (through DC-AC inverter): input 12 VDC, power consumption 190 W |
| Protection / ingress rating | Water resistance: IPX4 (without external power); Solids resistance: IP4X (brochure states compliance with the IP44 dust and water proof standard) |
| Drop resistance | Free fall complies with the requirements of 6.3.4.3, EN1789 (height of fall: 0.75 m); Bump complies with 6.3.4.2, EN1789 (medical devices for use in road ambulances); Shock and vibration complies with 21.102, ISO 9919 |
| Dimensions | Without external paddles: 295 mm (w) x 218 mm (d) x 279 mm (h); with external paddles: 295 mm (w) x 218 mm (d) x 323 mm (h) |
| Weight | Main unit: 5.5 kg; battery package (each): 0.57 kg; external paddle set: 0.86 kg |
| Regulatory / certifications | Safety: complies EN/IEC 60601-1; EMC: complies IEC 60601-1-2; AED sensitivity and specificity: meets AAMI DF-80; transport shock/vibration: ISO 9919; bump/free fall: EN1789 |
Our buying guide sets out how to compare defibrillator/monitor specifications; the models below cover different monitoring depths, display sizes and battery configurations.
Common questions about Mindray BeneHeart D6 Defibrillator / Monitor
How does the BeneHeart D6 differ from the D3?
The D6 has a larger 8.4-inch display with up to 4 waveform channels, adds 12-lead ECG, 2-channel IBP and 2-channel temperature monitoring, includes a built-in thermal recorder, and accepts two battery packs instead of one.
What defibrillation energy does the D6 deliver?
Manual mode uses 21 selectable settings from 1 J to 360 J with a biphasic truncated exponential waveform and impedance compensation. Charge time is under 3 s to 200 J and under 7 s to 360 J.
Which CO2 options are available?
Both Sidestream CO2 and Microstream CO2 are supported, alongside SpO2, NIBP, 3/5/12-lead ECG, 2-channel IBP and 2-channel temperature.
How long does the D6 run on battery?
With two 15.1 V / 5600 mAh lithium-ion packs the unit is rated for about 400 shocks at 360 J, 12 hours of monitoring or 9 hours of pacing.
Does the D6 integrate with hospital systems?
Yes. It transmits over 5G/2.4G Wi-Fi using IHE HL7 with DHCP, exports data via USB, and integrates with a central monitoring system and the Pre-Hospital Emergency Information System (PHEIS).
Can ICEN provide a quotation?
Yes. Send the model, quantity, destination market and required configuration through the Request a Quote button.
Need a quotation or product configuration?
Send the model, estimated quantity, destination market and required configuration - our export team will reply with availability, documentation and pricing.